Abstract

Kodak EDR2 film has been calibrated across the range of exposure conditions encountered in our cardiac catheterization laboratory. Its dose–response function has been successfully modelled, up to the saturation point of 1 Gy. The most important factor affecting film sensitivity is the use of beam filtration. Spectral filtration and kVp together account for a variation in dose per optical density of −10% to +25%, at 160 mGy. The use of a dynamic wedge filter may cause doses to be underestimated by up to 6%. The film is relatively insensitive to variations in batch, field size, exposure rate, time to processing and day-to-day fluctuations in processor performance. Overall uncertainty in the calibration is estimated to be −20% to +40%, at 160 mGy. However, the uncertainty increases at higher doses, as the curve saturates. Artefacts were seen on a number of films, due to faults in the light-proofing of the film packets.