Abstract
Gas analysis devices are measuring instruments and require periodic confirmation of their metrological parameters. This paper deals with methods of metrological support of measurements of absorption remote devices (with open cuvette). The design of the device where the operating and calibration cuvettes (with calibration gas mixture) are alternately (if necessary) placed on the optical path of the device, with the same optical elements. is proposed. This method of calibration and verification allows you to do without the use of CGM, which greatly simplifies calibration and verification of the gas analyzer at the workplace. In addition, the method reduces the calibration and verification error, since in the calibration mode the light flux passes along the same optical path as in the measurement mode, while the light flux and the degree of contamination of the optical elements by selective and non-selective components of the analyzed mixture remain unchanged.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Scientific journal of the Ternopil national technical university
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.