Abstract

The force-extension characteristics of orthodontic elastics made of natural rubber latex by 4 manufacturers were subjected to static testing under dry and wet conditions. The elastics consisted of 3 sizes: 3/16, 1/4, and 5/16 inch lumen sizes, each with forces specified according to the standard extension index of three times the lumen diameter. Most of the elastics did not match the specified index using the dry tests, but this should not be a serious clinical concern as all elastics showed acceptable regularity of force-extension characteristics. There was notable force degradation of all elastics when subject to water immersion, approximating 30% during the hour, but with an average less than 7% further loss up to 3 days. There were significant differences in force extension and force degradation characteristics between different extensions and force magnitudes for the elastics of the different manufacturers. It is suggested that the clinician could use the table of force degradation values for different extensions to select an appropriate elastic. (Am J Orthod Dentofacial Orthop 2000;118:280–7)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.