Abstract

In this paper, a relative calibration method is used to calibrate Fiber Bragg Grating(FBG) acceleration sensor. The calibrated sensor and standard sensor are fixed back-to-back to the excitation source. Both sensors receive the same excitation. The data from standard sensor and the FBG sensor are compared and analyzed. Data matching was performed by cross-correlation method and matching coefficients were calculated. The experimental results show that FBG vibration sensor is calibrated with a sensitivity of 1.3489 pm/g by cantilever beam vibration platform, and the correlation coefficient is 0.93. It has a good prospect in the field of acceleration sensing and can be widely used in actual acceleration measurement.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.