Abstract

<p>Two capacitance diaphragm gauges (CDGs) with 1333 Pa full scale, with a heated sensor head and an unheated one, respectively, were calibrated by three different methods; direct comparison to a resonant silicon gauge calibrated by a pressure balance, direct comparison to a CDG with 133 Pa full scale calibrated by a static expansion method, and the static expansion method. The calibration results of the three calibration methods show good agreement within their claimed uncertainties. Calibrated higher pressure points of CDGs by the pressure balance and lower pressure ones by the static expansion system are linearly interpolated within the calibrated uncertainty. Here, compensation of the thermal transpiration effect is important when a CDG is used with a heated sensor head.</p>

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.