Abstract

An alpha particle irradiator using 210Po radioactive sources was calibrated under specific conditions for cell irradiation. The energy spectrum of α-particles hitting the cell monolayer was measured by a surface barrier detector. The simulation of the results obtained with this detector was performed using the MCNPX Monte Carlo code with the same conditions used during the experimental measurements. Based on the experimental results the LET spectrum at the cell monolayer was extracted using the published Stopping Power and Range tables from the National Institute of Standards and Technology database. From this spectrum the average LET value was calculated to be (154 ± 9) keV/μm. In order to calculate the dose value at the cell monolayer, the alpha particle count rate hitting the cells was also estimated from the experimental measurements and compared with the activity of the source. Finally, the dose rate value at cell monolayer was calculated using the measured particle fluence.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.