Abstract

The results of calibration by cosmic muons of a shower lead-scintillation spectrometer of the sandwich type designed to work in high-intensity photon and electron beams with an energy of 0.1 - 1.0 GeV are presented. It was found that the relative energy resolution of the spectrometer depends on the angle of entry of cosmic muons into the spectrometer in the vertical plane and does not depend on the angle of entry in the horizontal plane. The relative energy resolution of the spectrometer was 16%. Placing an additional lead-scintillation assembly in front of the spectrometer improved the relative energy resolution of the spectrometer to 9%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.