Abstract
We present an in situ calibration technique for the LBNL positron emission mammography (PEM) detector module that is capable of measuring depth of interaction (DOI). The detector module consists of 64 LSO crystals coupled on one end to a single photomultiplier tube (PMT) and on the opposite end to a 64 pixel array of silicon photodiodes (PD). The PMT provides an accurate timing pulse, the PDs identify the crystal of interaction, the sum provides a total energy signal and the /spl Gamma/=PD/(PD+PMT) ratio determines the depth of interaction. We calibrate using the /sup 176/Lu natural background radiation of the LSO crystals. We determine the relative gain (K) of the PMT and PD by minimizing the asymmetry of the /spl Gamma/ distribution. We determine the depth dependence from the width of the /spl Gamma/ distribution with optimal K. The performance of calibrated detector modules is evaluated by averaging results from 12 modules. The energy resolution is a function of depth ranging from 24% FWHM at the PD end to 51% FWHM at the PMT end, and the DOI resolution ranges from 6 mm FWHM at the PD end to 11 mm FWHM at the PMT end.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.