Abstract

A newly developed neutron time-of-flight (nTOF) diagnostic with a fast instrument response function has been fielded on the OMEGA laser in a highly collimated line of sight. By using a small plastic scintillator volume, the detector provides a narrow instrument response of 1.7 ns full width at half maximum while maintaining a large signal-to-noise ratio for neutron yields between 1010 and 1014. The OMEGA hardware timing system is used along with an optical fiducial to provide an absolute nTOF measurement to an accuracy of ∼56 ps. The fast instrument response enables the accurate measurement of the primary deuterium-tritium neutron peak shape, while the optical fiducial allows for an absolute neutron energy measurement. The new detector measures the neutron mean energy with an uncertainty of ∼7 keV, corresponding to a hot-spot velocity projection uncertainty of ∼12 km/s. Evidence of bulk fluid motion in cryogenic targets is presented with measurements of the neutron energy spectrum.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.