Abstract

A modular photometric trap detector system has recently been developed at Physikalisch-Technische Bundesanstalt (PTB). All parts of the detector are now completely calibrated. The new planned traceability chain for the realisation of luminous intensity unit can therefore be established for the first time. This contribution shows the results of the individual calibration steps including the associated measurement uncertainties and correlations. A major part of the calibrations along the traceability chain is done at the upgraded measurement setup TULIP (TUnable Lasers In Photometry). The improvements of the TULIP setup are presented and the effects on the measurement uncertainty are shown. The result of the first complete calibration according to the new traceability chain is compared to previous calibration results both in terms of spectral irradiance responsivity and luminous responsivity. The further steps required towards implementing the new traceability chain and the possible implications are discussed.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call