Abstract

The HemoCue haemoglobin analyser consistently overestimated haemoglobin concentration ([Hb]) in the blood of all fish species (sockeye salmon Oncorhynchus nerka, Chinook salmon Oncorhynchus tshawytscha, Pacific bluefin tuna Thunnus orientalis and chub mackerel Scomber japonicus) by 22–50% (9·9–36·0 g l−1) over a [Hb] range of 20–160 g l−1. The systematic nature of the overestimation, however, allowed the formulation of an accurate calibration equation that can be used to correct values of [Hb] measured by the HemoCue in field studies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.