Abstract
Two reference methods have been established to calibrate the yield of a self-made 220Rn flow-through source. These reference methods include (a) a small scintillation cell flow-through method and (b) a gamma spectrometry method that takes advantage of an activated charcoal. The methods were validated theoretically and experimentally including intercomparison measurements. It is shown that the source is stable with a relative deviation less than 2.5% over 5 years, when the relative humidity ranged from 50 to 90%, and the temperature from 5 to 38 degrees C. The 220Rn monitor used in these reference methods is calibrated without a 220Rn gas source with determinate yield. On the basis of the calibration of the 220Rn source by the methods mentioned above, the accurate calibration of any 220Rn monitor by the self-made 220Rn source with stable yield is feasible.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.