Abstract
As a measurement system that can realize target detection and optical imaging, the accuracy of three-dimensional laser radar is a main performance index, which makes calibration an extremely important work. Traditional calibration methods have many disadvantages, such as harsh environment requirements, complex and tedious calibration processes, and inaccurate calibration results. To solve these problems, we propose a calibration method so that the relative position of the cooperative target and the detection sensor is fixed. The principle of space-time transform is used to simulate distance, and the synchronous control of distance is realized by controlling the delay module. In addition, a simple and practical calibration device is designed. In the actual measurement, the average absolute error is 0.0019 m, and the relative error is 0.0678% in the range of 0.5-25 m. The experimental results show that this method is stable and accurate, and it can calibrate the plane array laser radar quickly and accurately.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.