Abstract
This paper describes a calibration method for determining the physical location of the ultrasound (US) image plane relative to a rigidly attached 3D position sensor. A calibrated US probe can measure the 3D spatial location of anatomic structures relative to a global coordinate system. The calibration is performed by aiming the US probe at a calibration target containing a known point (1 mm diameter sphere) in physical space. This point is repeatedly collected at various locations in the US image plane to produce the calibration dataset. An idealized model of the collection process is used to eliminate outliers from the calibration dataset and also to examine the theoretical accuracy limits of this method. The results demonstrate accurate and robust calibration of the 3D spatial relationship between the US image plane and the 3D position sensor.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.