Abstract
A particle dosemeter (PD) is a payload of NEXTSat-2 in the low-earth orbit (LEO). The absorbed dose in LEO needs to be converted into the ambient dose equivalent (H*(10)). Due to a mixed field in LEO, the calibration factors (klow and khigh) should be determined for the low-and high-linear energy transfers (LET) (below and above 1.5keV/μm), respectively. The PD was irradiated with a 137Cs source at the Korea Radiation Solution facility to obtain H*(10) and absorbed doses. However due to the lack of sources for the high-LET calibration, H*(10) and an absorbed dose were calculated by simulating PD for the high-energy neutron field at CERN-EU high-energy Reference Field. The measured klow of PD had a difference of 5.1% and 9.5% from the calculated value of PD and the measured value of Liulin detectors, respectively. However, a difference in khigh between PD and Liulin was explained by the contribution of non-neutron components to Liulin in the measurements.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.