Abstract

A six-port reflectometer based on three coupled equal-mode-impedance microstrip lines of improved design was automated. An efficient calibration algorithm which allows the explicit determination of the calibration constants was used. The observed long-term stability of these constants suggested their determination and storage at all frequencies of interest. This drastically reduced the measurement time. Interpolation between the selected calibration frequencies is discussed, and the corresponding performance of the reflectometer is illustrated. The six-port results based on linear interpolation are found to be generally better and faster to obtain than those based on curve fitting using a cubic spline. The overall accuracy and the frequency range of operation can be increased by using a controllable generator with a higher output power.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">&gt;</ETX>

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.