Abstract
AbstractThe method of Balke, Hamielec, LeClair, and Pearce for GPC calibration and molecular weight computation is reviewed and evaluated. A new method, called GPCV2, is proposed which uses the single broad molecular weight distribution (MWD) standard for calibration developed by those workers, but which also incorporates the chromatographic dispersion σ to make the method more accurate in general use. Equations are presented for computing the relationship between actual MW and retention volume VR under the chromatographically broadened curves. This relationship is then compared to those obtained by using the Balke et al. method and the peak position method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.