Abstract

The National Institute of Standards and Technology (NIST), USA is participating in a project with the American Water Works Association Research Foundation (AwwaRF) to develop new guidelines for ultraviolet (UV) sensor characteristics to monitor the performance of UV water disinfection plants. The current UV water disinfection standards, ÖNORM M5873-1 and M5873-2 (Austria) and DVGW W294 3 (Germany), on the requirements for UV sensors for low-pressure mercury (LPM) and medium-pressure mercury (MPM) lamp systems have been studied. Additionally, the characteristics of various types of UV sensors from several different commercial vendors have been measured and analysed. This information will aid in the development of new guidelines to address issues such as sensor requirements, calibration methods, uncertainty and traceability. Practical problems were found in the calibration methods and evaluation of spectral responsivity requirements for sensors designed for MPM lamp systems. To solve the problems, NIST is proposing an alternative sensor calibration method for MPM lamp systems. A future calibration service is described for UV sensors intended for low- and medium-pressure mercury lamp systems used in water disinfection applications.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call