Abstract

Computer-generated hologram (CGH) is a diffractive optical element that has been commonly used as wavefront corrector in the interferometer and can be accurately manufactured to meet the requirements of high-precision detection. But the transmitted wavefront error (TWE) of the CGH substrate and the adjustment errors of cylindrical surface are influence factors of high accuracy measurement. In this paper, we propose the method of calibrating the CGH substrate TWE. After calibrating, the substrate TWE is 0.0052λ (RMS). Meanwhile, we obtain the general expression for the misalignment aberrations by describing the relationship between the wavefront of the measured cylindrical mirror and the corresponding adjustment errors. Then the adjustment errors of the test cylindrical surface can be completely separated by subtracting the fitting function from the measurement on the basis of least-square algorithm. After that, the PV value changes from 3.0565λ to 0.9825λ, and the RMS value changes from 0.5278λ to 0.1273λ, the measurement results have been greatly improved.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.