Abstract
We report on the calibration of a temperature stabilized ultra-low expansion (ULE) cavity using previously measured molecular tellurium and atomic cesium lines. By means of a dual frequency modulation technique, the frequency dependence of the free spectral range of the ULE cavity is measured and was found to vary by less than 60 Hz over the ∼55 THz range of the calibration. This method of calibration enables the ULE cavity to measure absolute frequencies to better than 1.5 MHz.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.