Abstract

We report on the calibration of a temperature stabilized ultra-low expansion (ULE) cavity using previously measured molecular tellurium and atomic cesium lines. By means of a dual frequency modulation technique, the frequency dependence of the free spectral range of the ULE cavity is measured and was found to vary by less than 60 Hz over the ∼55 THz range of the calibration. This method of calibration enables the ULE cavity to measure absolute frequencies to better than 1.5 MHz.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.