Abstract

The anisotropic resolution of linear arrays, tools that are widely used in diagnostics, can be overcome by compounding approaches. We investigated the ability of a recently developed calibration and a novel algorithm to determine the actual radial transducer array distance and its misalignment (tilt) with respect to the center of rotation in a 2-D and 3-D tomographic setup. By increasing the time-of-flight accuracy, we force in-phase summation during the reconstruction. Our setup is composed of a linear transducer and a rotation and translation axis enabling multidimensional imaging in ultrasound and photoacoustic mode. Our approach is validated on phantoms and young mice ex vivo. The results indicate that application of the proposed analytical calibration algorithms prevents image artifacts. The spatial resolution achieved was 160 and 250 μm in photoacoustic mode of 2-D and 3-D tomography, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.