Abstract

Considered the issues of X-ray dose control during diagnostic and therapeutic procedures using imaging tools. The dose of X-ray radiation from the visualization devices absorbed by the biological tissue of a person was determined when monitoring the position of the patient on the therapeutic table of the electron accelerator before the radiation therapy session. The processes of transmission of photons and electrons through the medium were simulated, and the X-ray spectra were measured. The emission spectrum of the Varian G-242 Rotating Anode X-ray Tube was obtained using an XR-100-CdTe spectrometer. The absorbed dose is calculated by the Monte Carlo method. The absorbed dose in the water phantom at tube voltage up to 80 kV was 0,9–1,5 mGy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.