Abstract
As one type of basic physical data related to the calculation of the construction of X-ray examination room shielding, we investigated the air-kerma rate 1 m from the source of the useful beam focus per unit current time product. Compared to the X-ray energy penetration values obtained in prior research, we obtained higher values. We found three causes for this discrepancy: (1) the X-ray tube total filtration (increasing total filtration reduces air kerma), (2) the tube voltage ripple percentage (increased voltage ripple reduces air kerma, and (3) the target angle (increasing the target angle increases the air kerma). On standardizing (1)-(3), we confirmed that our values mostly matched those obtained in past research. Consequently, when one employs the air-kerma rate that results from the consideration of (1)-(3), without pursuing a path of excessive safety, calculation of shielding under clinical conditions is possible.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.