Abstract

In the present work, we investigated the accuracy of the electron pencil‐beam redefinition algorithm (PBRA) in calculating central‐axis percent depth dose in water for rectangular fields. The PBRA energy correction factor C(E) was determined so that PBRA‐calculated percent depth dose best matched the percent depth dose measured in water. The hypothesis tested was that a method can be implemented into the PBRA that will enable the algorithm to calculate central‐axis percent depth dose in water at a 100‐cm source‐to‐surface distance (SSD) with an accuracy of 2% or 1‐mm distance to agreement for rectangular field sizes ≥2×2 cm. Preliminary investigations showed that C(E), determined using a single percent depth dose for a large field (that is, having side‐scatter equilibrium), was insufficient for the PBRA to accurately calculate percent depth dose for all square fields ≥2×2 cm. Therefore, two alternative methods for determining C(E) were investigated. In Method 1, C(E), modeled as a polynomial in energy, was determined by fitting the PBRA calculations to individual rectangular‐field percent depth doses. In Method 2, C(E) for square fields, described by a polynomial in both energy and side of square W [that is, C=C(E,W)], was determined by fitting the PBRA calculations to measured percent depth dose for a small number of square fields. Using the function C(E, W), C(E) for other square fields was determined, and C(E) for rectangular field sizes was determined using the geometric mean of C(E) for the two measured square fields of the dimension of the rectangle (square root method). Using both methods, PBRA calculations were evaluated by comparison with measured square‐field and derived rectangular‐field percent depth doses at 100‐cm SSD for the Siemens Primus radiotherapy accelerator equipped with a 25×25‐cm applicator at 10 MeV and 15 MeV. To improve the fit of C(E) and C(E, W) to the electron component of percent depth dose, it was necessary to modify the PBRA's photon depth dose model to include dose buildup. Results showed that, using both methods, the PBRA was able to predict percent depth dose within criteria for all square and rectangular fields. Results showed that second‐ or third‐order polynomials in energy (Methods 1 and 2) and in field size (Method 2) were typically required. Although the time for dose calculation using Method 1 is approximately twice that using Method 2, we recommend that Method 1 be used for clinical implementation of the PBRA because it is more accurate (most measured depth doses predicted within approximately 1%) and simpler to implement.PACS number: 87.53.Fs

Highlights

  • One of the practical goals for an electron-beam dose algorithm is that the algorithm be able to calculate dose delivered to the patient with an accuracy of 4% in low dose gradient regions or 2 mm distance to agreement (DTA) in high dose gradient regions.[1]

  • We investigated the accuracy of the electron pencil-beam redefinition algorithm (PBRA) in calculating central-axis percent depth dose in water for rectangular fields

  • PBRA calculations were evaluated by comparison with measured square-field and derived rectangular-field percent depth doses at 100-cm source-to-surface distance (SSD) for the Siemens Primus radiotherapy accelerator equipped with a 25×25-cm applicator at 10 MeV and 15 MeV

Read more

Summary

Introduction

One of the practical goals for an electron-beam dose algorithm is that the algorithm be able to calculate dose delivered to the patient with an accuracy of 4% in low dose gradient regions or 2 mm distance to agreement (DTA) in high dose gradient regions.[1] Boyd et al[2] showed that, with the inclusion of an incident energy spectrum, the pencil-beam redefinition algorithm (PBRA) of Shiu and Hogstrom[3] could meet this standard in water at a 100-cm source-tosurface distance (SSD) on a Varian Clinac 2100C (Varian Medical Systems, Palo Alto, CA). Boyd et al[4] showed that the addition of a dual-source model to account for collimationscattered electrons allowed the PBRA to meet this standard in water at 100-cm and 110-cm SSDs for selected energy-applicator combinations on a Varian Clinac 1800. Before any electron-dose algorithm—in this case the PBRA— can be judged acceptable for clinical implementation, the accuracy of calculated dose to water for all applicator, insert, and SSD combinations should be demonstrated for each beam energy commissioned

Objectives
Methods
Results
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.