Abstract
Both (125)I and (103)Pd sources have been widely used in the permanent prostate implant. An important consideration for the choice of brachytherapy sources is the relative biological effectiveness (RBE) for the source/seed used in the implantation. As RBE is closely related to the microdosimetric parameter, it is desirable to calculate the dose mean lineal energies for both (125)I and (103)Pd at various radial distances to the seed surface. Monte Carlo simulation was performed for photons emitted from (125)I and (103)Pd. Energy depositions from photons and all their secondary electrons were tracked. Dose distributions of lineal energy, d(y), were calculated for spheres of 1 microm in diameter and at various radial distances to the seed surface. From the dose distribution of lineal energy, the dose mean lineal energy, y(D), was derived. The results showed that the radiation qualities are constant in the distance range from 0.5 to 5 cm. In this distance range, the quality factor, relative to gamma rays from (60)Co, is 2.2 for (125)I and 2.5 for (103)Pd.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.