Abstract
Generalized cavity theory is used to calculate the dose–response of the radiochromic dye film MD55 to the X- and γ-radiations from 103Pd and 125I. The analysis indicates that for film calibrated using 60Co, the correction factor giving the dose to water is 1.81±0.05 for 103Pd and 1.71±0.05 for 125I.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.