We report a measurement procedure to determine simultaneously the major cation concentrations (Na, Ca, K and Mg) of seawater‐derived solutions by inductively coupled plasma‐atomic emission spectrometry. The best results were obtained when the IAPSO (‘standard’) seawater reference material was diluted by thirty times with Milli‐Q® water. We obtained an average reference value rK (the ratio of the mass fraction of potassium to that of chlorine, i.e., (g kg−1)/(g kg−1)) for IAPSO seawater of 0.0205 ± 0.0006 (2.9% RSD), not significantly different from 0.0206 ± 0.0005 (2.4% RSD) for seawater composition reported in the literature. The measured Na, Ca and Mg concentrations correspond to rNa, rCa and rMg values of 0.5406 ± 0.0026 (0.5% RSD), 0.02192 ± 0.00048 (2.2% RSD) and 0.06830 ± 0.00047 (0.7% RSD), respectively, in line with previous values measured by wet‐chemistry and atomic absorption spectrophotometry or wet‐chemical titration. Our measurement procedure was used successfully on synthetic seawater solutions and high‐temperature hydrothermal fluids.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.