Abstract

A new PVC-based membrane using p-isopropylcalix[6]arene (I) as an ionophore has been developed as a calcium-ion-selective sensor. Among various membranes prepared with and without plasticizer, the best performance was shown by a membrane with a composition (mg) I : NaTPB : PVC (2 : 2 : 120). This sensor exhibits a good potentiometric response to Ca2+ ion over a wide concentration range (3.9 × 10−6 to 1 × 10−1 M) with a near Nernstian slope (30 mV per decade of activity) and response time of 15 s. The sensor can be used for a period of 3 months without any drift in potential. The selectivity coefficient values are in the order of 1.0 × 10−3 for mono-, bi-, and trivalent cations, which indicates a good selectivity for Ca2+ ion. The useful pH range for the electrode was found to be 2.5–6.0, and it works well in mixtures with non-aqueous content up to 25% (v/v). The sensor can also be used successfully as an indicator electrode in the potentiometric titration of Ca2+ against EDTA.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.