Abstract

A colorimetric enzyme amplification system was used to develop an immunoassay for human calcitonin (hCT) with a sensitivity of 6 pmol/l, and intra- and inter-assay CVs of 12% and 11.8% respectively for the low pool, and 10% and 11.2% for the high pool. The mean recovery of added synthetic hCT (58.5 pmol) from the plasma of 10 patients was 110% (64.4 pmol). The correlation coefficient between radioimmunoassay (RIA) and amplified enzymo-immunoassay was found to be 0.96 (p 0.001). The assay was successfully applied to the measurement of elevated calcitonin levels in plasma from patients with medullary carcinoma of the thyroid (MCT). AEIA offered a reliable and sensitive alternative to RIA for calcitonin determination with the added advantage of convenience as the label employed was much more stable.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.