Abstract
In this study, the short-term reproducibility and agreement between individual units of dual X-ray and laser (DXL) Calscan (version 2), a new bone densitometry device based on dual X-ray absorptiometry and Laser technique, were evaluated. The variations in bone mineral density (BMD) and T-score between the right and the left foot were also studied retrospectively in 334 individuals. The short-term CV%, based on 19 healthy individuals scanned twice on 4 different units, was 2.1%, and the average CV% per scanner ranged from 1.1% to 1.8%. The short-term CV% in 33 elderly individuals (aged 83 ± 6 yr) was 3.9% and in a subgroup with a calcaneal T-score of −2.5 or below (n = 16) 5.1%. There was no statistical difference between the average BMD of the right and left foot, and ( r 2) between the right and left side was 0.84. The observed short- and long-term reproducibilities, as well as the acceptable variation between individual machines, show that this version of Demetech DXL Calscan is suitable for measuring BMD.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.