Abstract

In this study, the short-term reproducibility and agreement between individual units of dual X-ray and laser (DXL) Calscan (version 2), a new bone densitometry device based on dual X-ray absorptiometry and Laser technique, were evaluated. The variations in bone mineral density (BMD) and T-score between the right and the left foot were also studied retrospectively in 334 individuals. The short-term CV%, based on 19 healthy individuals scanned twice on 4 different units, was 2.1%, and the average CV% per scanner ranged from 1.1% to 1.8%. The short-term CV% in 33 elderly individuals (aged 83 ± 6 yr) was 3.9% and in a subgroup with a calcaneal T-score of −2.5 or below (n = 16) 5.1%. There was no statistical difference between the average BMD of the right and left foot, and ( r 2) between the right and left side was 0.84. The observed short- and long-term reproducibilities, as well as the acceptable variation between individual machines, show that this version of Demetech DXL Calscan is suitable for measuring BMD.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.