Abstract

A procedure for separation and enrichment of Cd(II) and Ni(II) ions based upon carrier element-free coprecipitation by using an organic coprecipitant, 2-{4-[2-(1H-Indol-3-yl)ethyl]-3-(4-methylbenzyl)-5-oxo-4,5-dihydro-1H-1,2,4-triazol-1-yl}-N′-(3-fluoro-phenylmethyliden) acetohydrazide, prior to their flame atomic absorption spectrometric detections has been developed. The effects of varied experimental conditions on the performance of the developed method such as pH, sample volume, amount of coprecipitating agent, etc. were evaluated in detail on the recovery of analyte ions, and the influences of some anions and cations were investigated. The limits of detection for Cd(II) and Ni(II) ions based on three times the standard deviation of the blanks (N: 10) were obtained as 0.70 μg L−1 and 1.21 μg L−1, respectively. The accuracy of the method was tested by analyzing a certified reference material and by spike tests. The method was applied to determine the levels of cadmium and nickel in stream and sea water, rice, red lentil, and wheat samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.