Abstract
The performance of liquid chromatography operation in open-tubular channels, the ideal chromatographic column format, is limited by slow mass transport between the mobile and stationary phase. We recently introduced a lateral mixing methodology ("vortex chromatography") to reduce Taylor-Aris dispersion by employing (small) AC-EOF (alternating current electroosmotic flow) fields oriented perpendicular to the conventionally applied, axially oriented pressure gradient, resulting in the reduction of the C-term by a factor of 3, studied in 40 × 20 μm2 (aspect ratio (AR) = 2) channels under unretained conditions. In the present contribution, a further increased performance gain for channel dimensions relevant for chromatographic applications is demonstrated. The impact of the applied voltage and salt concentration is studied for 3 × 20 and 5 × 20 μm2 channels in ARs of up to 6.7, revealing a C-term reduction potential of a factor of up to 5 for large molecules (dextran) under unretained conditions. The decrease in κaris in a 5 μm channel (reduction of 80%) was larger than the decrease in a 3 μm channel (reduction of 44%).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.