Abstract
In this article we describe the setup, installation, calibration, and operation of the two-color (CO2–HeNe) laser heterodyne interferometer installed in the TJ-II Stellarator. This diagnostic has been routinely working during the last TJ-II experimental campaign (Autumn 2003) in low-density [electron cyclotron resonance (ECRH) heated, ne≅1019 m−3] and medium density [neutral beam injection (NBI) heated, ne≅5×1019 m−3] plasmas with line integral error better than 2×1017 m−2. This work demonstrates the possibility of using low-cost two wavelength laser interferometric diagnostics to measure low and medium-density plasmas instead of the complicated far infrared (FIR) interferometric systems that also suffer from fringe-jumps errors. The issues of low sensitivity to plasma density and mechanical vibration subtraction typical of these laser interferometers are solved through proper calibration of the interferometric system in the machine. First results on NBI heated plasmas measured by the laser interferometric system in the Stellarator TJ-II will be shown.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.