Abstract

Ultrasound detection via optical resonators can achieve high levels of miniaturization and sensitivity as compared to piezoelectric detectors, but its scale-up from a single detector to an array is highly challenging. While the use of wideband sources may enable parallel interrogation of multiple resonators, it comes at the cost of reduction in the optical power, and ultimately in sensitivity, per channel. In this work we have developed a new interferometric approach to overcome this signal loss by using high-power bursts that are synchronized with the time window in which ultrasound detection is performed. Each burst is composed of a train of low-noise optical pulses which are sufficiently wideband to interrogate an array of resonators with non-overlapping spectra. We demonstrate our method, termed burst-mode pulse interferometry, for interrogating a single resonator in which the optical power was reduced to emulate the power loss per channel that occurs in parallel interrogation of 20 to 200 resonators. The use of bursts has led to up 25-fold improvement in sensitivity without affecting the shape of the acoustic signals, potentially enabling parallel low-noise interrogation of resonator arrays with a single source.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call