Abstract

AIM: To report problems in the tethering catheter shaft of the Tempofilter, temporary caval filter.MATERIALS AND METHODS: Two cases are reported where the tethering catheter shaft of the Tempofilter buckled within the jugular vein.RESULTS: Buckling of the tethering shaft caused cephalic migration of the filter into the right atrium. Both filters were uneventfully removed without adverse sequelae.CONCLUSION: These cases demonstrate another mechanism of tethering shaft shortening which results in unexpected migration of the filter. This device is potentially dangerous because of the liability of the tethering cather to buckle at or near the insertion site with cephalic migration of the filter. Improvements in the tethering catheter mechanism need to be made before further patient usage. Kelly, I.M.G. and Boyd, C.S. (1999) Clinical Radiology54 , 398–401.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.