Abstract

Two years after a quadrupled hamstring anterior cruciate ligament reconstruction using polylactic acid cross pin femoral fixation, a 32-year-old woman presented with symptoms of knee catching, locking, and stiffness. Diagnostic arthroscopy revealed a loose body in the anterior compartment of the knee which was determined to be part of the polylactic acid femoral fixation pin. The graft was intact and well fixed. After simple arthroscopic removal, the patient returned to full activities and resumed normal function postoperatively. We speculate that the fixation pins may have entered the notch and later degraded or fractured. Using axial magnetic resonance images, we provide preliminary data suggesting that pins angled posterior to the epicondylar axis may violate the notch. If pins are to be placed posterior to the epicondylar axis, maximum pin length can be estimated by the formula: 0.4 x the interepicondylar distance.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.