Abstract
BackgroundEpidural anaesthesia is one of the most used neuraxial anaesthesia techniques. It has been utilized as the principal anaesthesia modality involving lower limb surgery. Breakage of epidural catheters is an unusual occurrence whose subsequent therapy lacks uniformity due to the absence of a consensus and defined standards.Case presentationA 39-year-old male with no comorbidities was scheduled for right lower limb reconstruction surgery due to non-union of the tibia. Combined spinal-epidural anaesthesia was planned. During epidural catheter insertion, there was difficulty threading the catheter, and upon its removal, a long segment of catheter (8 cm) was left inside the body. Following informed consent from the patient, the retained fragment was surgically extracted in the same sitting under general anaesthesia.ConclusionsEven when faced with such a circumstance, it is best to explore surgically and remove a long segment broken catheter to allay patient anxiety.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.