Abstract

This paper describes the principle of a new fully automated digitally assisted coaxial bridge having a large bandwidth ranging from 60 Hz to 50 kHz. The performance of the bridge is evaluated making comparisons between calculable ac resistors. The agreement between the calculated and the measured frequency dependence of the resistors is better than at frequencies up to 5 kHz, better than up to 20 kHz and better than up to 50 kHz. This bridge is particularly well suited to investigate the ac transport properties of graphene in the quantum Hall regime.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.