Highlights Abstract Objective: To reduce the incidence of central line-associated bloodstream infection (CLABSI) in peripherally inserted central catheters (PICC) through the introduction of an antimicrobial (AM) catheter as recommended in evidence-based guidelines and standards. Design: Quality improvement project comparing incidence of infections pre-implementation and postimplementation of the new catheter. Setting: A 582-bed community teaching hospital in Northwest Indiana. Methods: Pre-implementation analysis of surveillance data indicated that 50% of CLABSIs occurred in patients with PICCs in situ. A gap analysis was performed to review institutional practices against evidence-based recommendations. The use of an AM catheter was supported in each of the documents consulted. After introduction of the new device, performance was measured in a prospective manner using standardized Centers for Disease Control and Prevention (CDC) National Healthcare Safety Network (NHSN) surveillance protocols for CLABSI and internal data sources for other measures. Results: After 30 months of data collection, the PICC CLABSI incidence reduced from a baseline rate of 1.83/1000 PICC days to 0.162/1000 PICC days (91.15% reduction, P = 0.0002). Conclusion: Combined with continued compliance with basic prevention strategies (i.e., use of a central line insertion checklist/insertion bundle) and optimization of device selection and lumen justification, the introduction of an antimicrobial/antithrombogenic (AM/AT) PICC was associated with a significant reduction in CLABSI.