Abstract

Many medical linear accelerators generate not only high-energy photons, but also high-energy electrons, and they are no longer equipped with beam stoppers. Therefore, shields might be necessary against bremsstrahlung and photoneutron generated by high-energy electron beams. However, there are few physical studies, and no recommendations are made about shields nowadays. In this report, the leakage doses of bremsstrahlung and photoneutron were calculated by the use of Monte Carlo simulation. To verify the calculated results, the photoneutron leakage dose was measured with a rem counter. The results clearly show that the bremsstrahlung and photoneutron leakage dose generated by electron beams of 24 MeV or below is negligible.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.