Abstract

In continuous positive airway pressure(CPAP) devices, pressure can be generated by two different mechanisms: either via an expiratory valve or by one or more jets. Valved CPAP devices are referred to as constant-flow devices, and jet devices are called variable-flow devices. Constant-flow CPAP devices are said to reduce the imposed work of breathing due to lower breath-dependent pressure fluctuations. The present study investigates the performance of various constant-and variable-flow CPAP devices in relation to breath-dependent pressure fluctuations. Experimental study comparing the pressure fluctuations incurred by seven neonatal CPAP devices attached to an active neonatal lung model. Spontaneous breathing was simulated using a tidal volume of 6 ml at pressure levels of 5, 7, and 9 mbar. The main outcomes were respiratory pressure fluctuations, tidal volume, and end-expiratory pressure. All CPAP devices tested showed respiratory pressure fluctuations, varying from 0.631to 3.466 mbar. The generated tidal volume correlated significantly with the pressure fluctuations (r = -0.947; p = 0.001) and varied between 5.550and 6.316 ml. CPAP devices with jets showed no advantage over CPAP devices with expiratory valves. End-expiratory pressure in the nose deviated from the set pressure between -1.305and 0.644 mbar and varied depending on whether the pressure was measured in the device or in the tube extending to the nose. During standard spontaneous breathing, breath-dependent pressure fluctuations in constant-and variable-flow devices are comparable. Pressure measurements taken in the tubing system can lead to a considerable deviation of the applied pressure.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call