Abstract
Herein, we describe an unusual case of cystoscope damage during a planned laser cystolithotripsy in a 65-year-old male with a previous history of radical prostatectomy for prostate cancer and subsequent serial urethral dilations for bladder neck contracture. Upon crossing the penile urethra without exerting significant pressure, we noticed the cystoscope's distal metallic tip detachment. Therefore, we re-introduced another 22Fr cystoscope and removed the broken part with alligator forceps. Fortunately, no urethral injury or associated complications were noticed on gently re-entering the bladder. Hence, we managed to complete the endoscopic laser cystolithotripsy shortly thereafter. Review of the relevant literature revealed three similar cases. All related to the same manufacturer. Urologists should not lose sight of the fact that such an unexpected instance may tremendously impact the procedure's success, requiring vigilance and adherence to safety protocols.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.