Abstract

Infections related to deep brain stimulation (DBS) devices are not rare, but abscess formation in brain parenchyma is extremely rare. A 50-year-old man with generalized dystonia had undergone DBS of bilateral globus pallidus internus. The authors attempted to remove the bilateral DBS system due to repeated device infections caused by metal allergies. However, the intracranial lead had to be left in place, because the lead was strongly adherent to brain parenchyma. Five years later, magnetic resonance imaging showed ring-like enhancement localized around the tip of the intracranial lead, suggesting brain abscess. In response to the symptoms, the remaining left intracranial electrode was removed. Brain abscesses require several months of treatment with appropriate antibiotics, but good outcomes can be achieved with appropriate treatment. Brain abscess is a rare complication of DBS. In the present case, the infection spread from the subcutaneous infected foci to the intracranial area through the lead, resulting in the formation of a brain abscess. Removing as much of the device as possible from the body is therefore important, even if adhesions with brain parenchyma or other tissues are present, because of the risk of serious complications, as seen in this case.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.