Abstract
A prototype in-body gamma camera system with integrated trans-rectal ultrasound (TRUS) and associated real-time image acquisition and analysis software was developed for intraoperative source tracking in high dose rate (HDR) brachytherapy. The accuracy and temporal resolution of the system was validated experimentally using a deformable tissue-equivalent prostate gel phantom and a full clinical HDR treatment plan. The BrachyView system was able to measure 78% of the 200 source positions with an accuracy of better than 1 mm. A minimum acquisition time of 0.28 s/frame was required to achieve this accuracy, restricting dwell times to a minimum of 0.3 s. Additionally, the performance of the BrachyView-TRUS fusion probe for mapping the spatial location of the tracked source within the prostate volume was evaluated. A global coordinate system was defined by scanning the phantom with the probe in situ using a CT scanner, and was subsequently used for co-registration of the BrachyView and TRUS fields of view (FoVs). TRUS imaging was used to segment the prostate volume and reconstruct it into a three-dimensional (3D) image. Fusion of the estimated source locations with the 3D prostate image was performed using integrated 3D visualisation software. HDR BrachyView is demonstrated to be a valuable tool for intraoperative source tracking in HDR brachytherapy, capable of resolving source dwell locations relative to the prostate anatomy when combined with TRUS.