Abstract

Abstracts Concave eye applicators with 90Sr/90Y and 106Ru/106Rh beta-ray sources are usually used in brachytherapy for the treatment of superficial intraocular tumors as uveal melanoma with thickness up to 5 mm. The aim of this work consisted in using the Monte Carlo code MCNPX to calculate the 3D dose distribution on a mathematical model of the human eye, considering 90Sr/90Y and 160Ru/160Rh beta-ray eye applicators, in order to treat a posterior uveal melanoma with a thickness 3.8 mm from the choroid surface. Mathematical models were developed for the two ophthalmic applicators, CGD produced by BEBIG Company and SIA.6 produced by the Amersham Company, with activities 1 mCi and 4.23 mCi respectively. They have a concave form. These applicators' mathematical models were attached to the eye model and the dose distributions were calculated using the MCNPX ⁎F8 tally. The average doses rates were determined in all regions of the eye model. The *F8 tally results showed that the deposited energy due to the applicator with the radionuclide 106Ru/106Rh is higher in all eye regions, including tumor. However the average dose rate in the tumor region is higher for the applicator with 90Sr/90Y, due to its high activity. Due to the dosimetric characteristics of these applicators, the PDD value for 3 mm water is 73% for the 106Ru/106Rh applicator and 60% for 90Sr/90Y applicator. For a better choice of the applicator type and radionuclide it is important to know the thickness of the tumor and its location.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call