Abstract

A three-dimensional projection reconstruction technique is described for imaging boron-11 distributions, with potential application to boron neutron capture therapy. The method samples a spherical volume of k space uniformly to obtain a 32 x 32 x 32 matrix with voxel size of 0.42 cm3. A signal-to-noise ratio (S/N) of 3 was obtained in 8.5 minutes in a phantom containing 75 micrograms/mL of boron in borocaptate sodium (BSH). Images were obtained in a dog after cessation of an intravenous infusion of BSH and again 30 minutes later, with a maximum boron S/N of about 12. Boron levels in the brain dropped about 6%-8% and were more diffusely distributed on the images obtained 30 minutes after BSH infusion.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.