Abstract
A BC523A liquid scintillator loaded with boron-10 was tested as a detector for both fast and thermal neutrons. A pulse shape discrimination (PSD) method based on a zero-crossing principle was applied to distinguish between neutron and gamma radiation. High quantum efficiency Photonis XP5500B photomultiplier was used to enhance light detection from the scintillator. This allowed a good registration of the energy spectrum of neutron capture events on boron-10, corresponding to about 60 keV gamma-rays. The applied PSD method proved to be useful for n/gamma discrimination. A good resolving power of the method was achieved even without gating on neutron capture events. A comparison with a standard BC501A liquid scintillator was done to evaluate thermal neutrons detection efficiency in BC523A.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.