A new technique to biomimetic deposition calcium phosphate (BDCaP) coating onto rough titanium substrates has been developed recently. This biomimetic deposition technique seems to be promising. It appears to have some advantages such as an inexpensive and simple set-up, and the possibility to synthesize layers with a defined surface morphology. The aim of this study was to examine the bone responses to BDCaP-coated implants in a rabbit model. Thirty one implants (16 BDCaP and 15 rough) were inserted into both tibia of 15 rabbits. After 2, 4, and 8 weeks following the implantation, the tibias were retrieved and prepared for histological evaluation. After 2 weeks, BDCaP-coated implant showed more bone ingrowth inside threads than the rough implants in medullary region (31.43% vs. 24.38%). Histological and quantitative histomorphometrical measurements demonstrated no more bone ingrowth and bone-implant contact for coated implant as compared with uncoated implant in cortical region at all experiment periods. From the histological viewpoint, the BDCaP coating did not have any positive effect on new bone formation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.