Failure rates at second-stage surgery were reported for the ongoing Dental Implant Clinical Research Group studies of the Spectra-System (Core-Vent Corporation, Las Vegas, NV) implants. As of May 1995, 69 implants failed out of 2,633 placed and uncovered. The overall failure rate was 2.6%, with 3.6% in bone quality 1 (BQ-1), 2.4% in BQ-2, 2.5% in BQ-3, and 3.1% in BQ-4. HA-coated cylinders had the lowest number of failures and titanium alloy baskets the highest. The basket design failed more often in the posterior jaw areas; 9 of 32 clinical centers (28%) accounted for 72% of these failures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.