Abstract

Bone lead serves as a better, more accessible biomarker to many communities experiencing chronic exposure to lead. A new method using low energy x-ray fluorescence in a handheld device (portable XRF) allows us to measure this chronic biomarker in only a few minutes. However, many unknowns remain about this biomarker measured using a new low energy x-ray technique. The low energy of the new method was theorized to measure a slightly different portion of the bone than previous techniques, which could influence measurements at different bone sites and types. We tested how bone measurements varied across five bone sites: mid-tibial shaft, proximal tibia, distal tibia (ankle), ilium, and cranium. We found bone lead measurements are not significantly different between skeletal elements when measured using a portable XRF. On average, bone lead in the repository samples was measured to be 21.6 ± 21.3 μg/g with an XRF detection limit of 2.1 ± 0.5 μg/g. Cumulative lead exposure can be effectively measured using the portable XRF on a variety of bone types, but the tibia should be preferentially measured to compare between studies and individuals.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.